Vocalization removal for improved automatic segmentation of dual-axis swallowing accelerometry signals

September 1, 2011

Automatic segmentation of dual-axis swallowing accelerometry signals can be severely affected by strong vocalizations. In this paper, a method based on periodicity detection is proposed to detect and remove such vocalizations. Periodic signal components are detected using conventional speech processing techniques and information from both axes are combined to improve vocalization detection accuracy. Experiments with 408 healthy subjects performing dry, wet, and wet chin tuck swallows show that the proposed method attains an average 95.3% sensitivity and 96.3% specificity. When applied in conjunction with an automatic segmentation algorithm, it is observed that segmentation accuracy improves by approximately 55%. These results encourage further development of medical devices for the detection of swallowing difficulties.

DOI: 10.1016/j.medengphy.2010.04.008

This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author’s copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.